Archives of Medical Case Reports
ISSN: 2691-7971

Case Report - Archives of Medical Case Reports (2021) Volume 3, Issue 1

Pseudothrombocytosis Due to Red Blood Cell Fragmentation in Haemoglobin-H Disease

Ahmed Maseh Haidary1*, Raja Zahratul Azma2, Sarah Noor3, Maryam Ahmad1

1Department of Pathology and Clinical laboratory, French Medical Institute for Mothers and Children, Kabul Afghanistan

2Department of Pathology and Laboratory Medicine, The National University of Malaysia, Kuala Lumpur Malaysia

3Department of Haemato-Oncology, Jumhoriat Hospital, Kabul Afghanistan

*Corresponding Author:
Ahmed Maseh Haidary
E-mail:ahmedmaseh.haidary@fmic.org.af

Received date: September 14, 2021; Accepted date: September 22, 2021

Citation: Haidary AM, Azma RZ, Noor S, Ahmad M. Pseudothrombocytosis Due to Red Blood Cell Fragmentation in Haemoglobin-H Disease. Arch Med Case Rep. 2021; 3(1): 29-32.

Copyright: © 2021 Haidary AM, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.


Abstract

Introduction: Red cell fragmentation may result in erroneously high platelet counts by automated blood counters. Therefore, any abnormal platelet counts in a full blood count analysis should be re-evaluated either using blood film manual counting or with more accurate automated analysis options. Case Report: We report a case of Haemoglobin-H disease that was noted to have a very high platelet count, by newly installed automated blood analyser, during regular follow up. Manual platelet counts and the automated analysis using flowcytometric counting of the same analyser showed a normal platelet count. Conclusion: A wide spectrum of blood disorders can cause pseudothrombocytosis. Whenever an unexpected high platelet count is encountered, alternative modalities should be used to confirm the actual platelet count such as manual count on a blood film. Furthermore, machines using the impedance principle need to be calibrated to reduce the chances of getting abnormal results with actual normal platelet counts.


Keywords

Pseudothrombocytosis, Red cell fragmentation, Hb-H disease


Abbreviations

Hb-H: Haemoglobin-H; RNA: Ribonucleic Acid; RBC: Red Blood Cells; BA: Basophil; EO: Eosinophil; HCT: Haematocrit; HGB: Haemoglobin; LY: Lymphocytes; MCH: Mean Corpuscular Haemoglobin; MCHC: Mean Corpuscular Haemoglobin Concentration; MCV: Mean Corpuscular Volume; MO: Monocytes; NE: Neutrophils; NRBC: Nucleated Red Blood Cells; PLT: Platelets; RDW: Red Cell Distribution Width; WBC: White Blood Cells; pBA: Percentage of Basophils; pEO: Percentage of Eosinophils; pLY: Percentage of Lymphocytes; pMO: Percentage of Monocytes; pNE: Percentage of Neutrophils; pNRBCs: Percentage of Nucleated Red Blood Cells


Introduction

Red cell membrane in patients with haemoglobin H (Hb-H) disease and other haemoglobinopathies is rigid and thus red blood cells are fragmented during passage through narrow capillary beds, especially through the splenic sinusoids and the liver [1,2]. These fragments are frequently seen in peripheral blood films [2].

Automated blood cell counting systems, if not adequately calibrated then may report these red cell fragments as platelets. This pseudothrombocytosis can be confirmed either with manual platelet counting on a peripheral blood smear or using flow-cytometry mode, which is considered to be the gold standard for correct estimation of platelet count that uses RNA fluorescent dyes to count the true platelet values [3]. Here, we report a patient with Hb-H disease who exhibited an unusually high platelet count.


Case Presentation

A 59-year-old lady with non-transfusion dependent Hb-H disease on regular follow up, was noted to have a very high platelet count, highest recorded value being 2158 × 109/L, during her clinic visits, as shown in Figure 1. Two months earlier the platelet count was 339 × 109/L, also shown in (Figure 1). The patient had no signs or symptoms related to hyper-viscosity, or any bleeding tendencies. On examination the patient had no hepatosplenomegaly. She was subsequently followed up at the haematology clinic 6 months apart. Two subsequent FBCs also showed platelet counts more than 2000 × 109/L. All the high values were obtained using the Impedance mode by a newly installed Sysmex XN3000, shown in Figure 1. A full blood picture was requested and showed aniso-poikilocytosis with numerous red blood cell fragmentation, with manual platelet count of 60-80 platelets per high power field (approximately equivalent to 300-400 × 109/L), as shown in (Figure 2). Finally, the FBC was repeated using the Sysmex XN3000 via optical mode with reticulocyte-platelet channel that showed a normal platelet count of 153 × 109/L, as shown in Figure 1.


Discussion

A conventional haematology analyzer uses two main principles for the estimation of an absolute blood cell count [4]. The first and oldest principle is the “Coulter’s impedance principle”. According to Coulter impedance principle, in a closed electrical circuit through a conducting fluid medium, an impedance is created when small particles such as blood cells pass through an aperture between the two electrodes. The impedance to the flow of charges between two electrodes is equivalent to the size of the particle, as shown in (Figure 3) [5]. The computer in a well calibrated analyzer will then count each impedance event as either red blood cell, leukocyte or platelet depending on the size of the particle. Working on the impedance principle, the analyzer needs to be well calibrated to avoid erroneous measurement of the counts, as shown in this case. Frequent troubleshooting of abnormal results and thus frequent calibration is the key to achieving best performance of a modern multiparameteric haematology analyzer [6,7].

The second principle is based on the principle of flowcytometry designated to optical reticulocyte and platelet Reticulocyte/platelet channels [8]. A Fluorescent dye that specifically binds to RNA component of the cells is used in this method [8]. It should be recalled that reticulocytes and platelets contain excess amounts of RNA material in their cytoplasm, the Reticulocyte/platelet channel is particularly specific and sensitive for accurate estimation of reticulocyte as well as platelet counts [8]. This is also a useful method to assess the immature platelet fraction, in patients with increased peripheral destruction of platelets [9].

In our patient the cause of pseudothrombocytosis was the red cell fragments, produced as a result of underlying Haemoglobin-H (Hb-H) disease. A review of the literature showed that pseudothrombocytosis is caused by a wide range of disorders such as microerythrocytosis due do iron deficiency anaemia [3,10], Haemoglobin H disease with red cell fragmentation [11], cryoglobulinaemia [12] and various red cell fragmentation syndromes including microangiopathic haemolysis [3,13].


Conclusion

In conclusion, laboratory personnel should be aware that analyzers that use the impedance principle might lead to spurious blood counts in a wide spectrum of blood disorders. When an unexpectedly high platelet count is encountered, the count should be confirmed using more accurate modalities, including manual count on a blood smear. Haematologic analyzers using impedance mode need to be properly calibrated to reduce the chances of getting abnormal results with normal actual counts.


Authors’ Contributions

AMH and SRN conceived the idea. SRN and AMH were the major contributor to the writing of the manuscript. AMH, RZA diagnosed the case. MA collected the data from hospital integrated laboratory management system (ILMS), AMH and SRN were the major contributors for critically revising the manuscript for important intellectual content. SRN and RZA have given expert opinion and final approval of the version to be published. All authors read and approved the final manuscript.


Funding

Authors received no funding for the writing.


Availability of Data and Materials

All the generated data are included in this article.


Ethical Approval and Consent to Participate

Not applicable.


Consent for Publication

Written informed consent was obtained from patient’s legal guardian for publication of this case report and all the accompanying figures. A copy of the written consent shall be availed to the Editor-in-Chief of this journal.


Competing Interests

The Authors declare that they have no competing interests.


References